目錄:上海鉑勒機(jī)電設(shè)備有限公司>>HACH在線水質(zhì)分析儀>>在線電極及其他附件>> D3422C3
參考價(jià) | 面議 |
參考價(jià) | 面議 |
更新時(shí)間:2016-10-26 12:48:40瀏覽次數(shù):685評(píng)價(jià)
聯(lián)系我們時(shí)請(qǐng)說(shuō)明是化工儀器網(wǎng)上看到的信息,謝謝!
These enhanced performance sensors are manufactured to exacting tolerances using high quality, rugged materials for demanding applications including ultra-pure water, clean-in-place (CIP), and boiler/condensate monitoring. Each sensor is tested to determine its unique, absolute four-digit cell constant. Simply key in this constant (Hach’s easy DRY-CAL™ method) when configuring the analyzer to ensure the highest possible measuring accuracy. Also, each sensor has a Pt 1000 RTD temperature element built into its tip for exceptionally fast response to changes in temperature with ±0.1ºC accuracy.
These enhanced performance sensors measure from theoretically pure water (0.057 µS/cm or 18.2 MΩ) up to 200,000 µS/cm. Hach’s sc Digital Controllers accept multiple digital sensor inputs, and can be user-set to measure conductivity, resistivity,TDS, salinity, or one of six calculated measurements.
Compression fitting sensors—Feature titanium electrodes and a compression fitting for universal installation with up to 4 inches (102 mm) insertion depth. The 1/2-inch or 3/4-inch male NPT compression fittings are offered in Kynar® (PVDF) or 316 stainless steel. A longer version of this sensor is available for use with a 316 stainless steel ball valve hardware assembly to insert/retract the sensor from the process without stopping the flow. The longer version can also be used for insertion through a compression fitting. Maximum insertion depth is 7 inches (178 mm).
There are no complicated wiring or set up procedures with any Hach sc controller. Just plug in any combination of Hach digital sensors and it's ready to use—it’s “plug and play.”
Includes: sensor with cable and manual
Accuracy:± 2 % of reading above 200 µS/cm
Cable Length:Standard Probe Cable Length (integral)Digital Probe: 7 m (23 ft.)
Operating Temperature Range:-20 to 200 °C
Pressure Limit:0 to 300 psi (20.7 BAR)
Repeatability:± of reading
Response Time:90 % of reading within 30 seconds of step change
Temperature Compensation:Temperature Compensator: Pt 1000 RTD
Temperature Measurement:-20 to 200 °C (–4 to 392°F)
Transmission Distance:1000 m (3280 ft.), maximum when used with a termination box