目錄:上海鉑勒機電設備有限公司>>HACH在線水質(zhì)分析儀>>在線電極及其他附件>> 5790001
An internal standard calibrates the instrument with every reading. Factory calibration and membrane replacement are unnecessary. Periodic wiping of the sensor with a wet rag is all that is required for continuous measurement.
The sensor is unaffected by pH swings, hydrogen sulfide, wastewater chemicals, heavy metals, or organic build-up on the sensor. Warm-up time is unnecessary so the analyzer can start measuring within 30 seconds of when it’s turned on.
The HACH LDO probe can be installed with a choice of pole or ball-float mounting kits. The probe has only one inexpensive replacement part—the sensor cap that is simple to replace. The sensor cap is warranted for one year. The probe is warranted for three years.
The HACH LDO sensor is coated with a luminescent material. Blue light from an LED is transmitted to the sensor surface. The blue light excites the luminescent material. As the material relaxes it emits red light. The time for the red light to be emitted is measured. Between the flashes of
blue light, a red LED is flashed on the sensor and used as an internal reference.
Increased oxygen in the sample decreases the time it takes for the red light to be emitted. The time measurements correlate to the oxygen concentration.
There’s no complicated wiring or set up procedures with the Hach sc100 controller. Just plug in any Hach digital sensor and it’s ready to use—it’s “plug and play.”
One or two sensors—Use the sc100 Digital Controller to receive data from up to two Hach dig
Accuracy:Below 1 ppm:
± 0.1 ppm Below 1 ppm
Accuracy 2:± 0.2 ppm Above 1 ppm
Accuracy 3:Temperature:
± 0.2 °C
Hazardous Location Ratings:ETL listed (cETLus marked) to Canadian and US General Safety and Hazardous (Class l, Div. 2) Locations
Measurement range:0 to 200 % saturation
Measuring Range:to
Operating Temperature Range:0 to 50 °C
Pressure Limit:max
Repeatability:± of span
Resolution:± Below 10 ppm 1% saturation
Response Time:at 20ºC: To 95 % in less than 60 seconds
Response Time - 20 °C:To 90 % in less than 40 seconds
Sensor Immersion Depth:Pressure Limits at 107 m 1050 kPa (350 ft.), 1050 kPa (150 psi), maximum
Sensor Thread:Inner thread
Storage Conditions:-20 °C to 70 °C
Transmission Distance:1000 m (3280 ft.) maximum when used with a termination box
Warranty:3 years
Weight:3.09 lbs. (1.4 kg)