目錄:上海鉑勒機電設備有限公司>>HACH在線水質分析儀>>在線電極及其他附件>> LXV441.99.11302
Even the smallest oil traces can impair water quality. The FP 360 sc is specifically designed to detect traces of mineral oils in water while providing the necessary value and benefits for a positive return on investment. Monitors surface water, process water and industrial water continuously. Has submersible probe design. No tubes, pumps or valves - simply wipe off the sensor's measurement window. Calibrate once every two years. Available in stainless steel or titanium.
The FP 360 sc is specifically designed to detect traces of mineral oils in water while providing the necessary value and benefits for a positive return on investment.
Due to its unique combination of submersible probe design and UV fluorescence sensing technology, the FP 360 sc delivers the best technology to detect oil in water and is priced below competitive UV fluorescent instruments.
The FP 360 sc has no tubes, pumps, or valves that can foul or require constant maintenance interventions. Maintenance is limited to occasional wiping of the sensor's measurement window, calibration once every two years, and Xenon lamp replacement every four years.
While laboratory testing is the ultimate method of measuring oil in water, it is a long and complex process that requires special equipment and trained lab personnel. The FP 360 sc provides a cost-effective, continuous online monitoring solution to maintain process control and avoid oil contamination with minimal laboratory testing.
The FP 360 sc can detect and measure polycyclic aromatic hydrocarbons (PAHs) from 1.2 ppb to up to 5000 ppb (µg/L). This is approximay equivalent to a concentration of mineral oil between 0.1 to 150 ppm (mg/L). Furthermore, the FP 360 sc method of detection makes it impervious to interferences by turbid water or natural organic and biological matter that impact online light scattering, UV absorbance, and VIS fluorescence instruments.
Oil-in-water probe, user manual
Ambient Temperature:-5 °C to 45 °C (23 °F to 113 °F)
Cable Length:1.5 m (4.9 ft) - 40 m (130 ft) max
Calibration:Factory calibrated with UV fluorescence standard or process calibration with results of a grab sample analysis.
Detector:UV photodiode with interference filter; Compensation of daylight and flashlamp intensity fluctuations
Dimensions (D x L):2.68 in x 12.24 in (68 mm x 311 mm)
Excitation:254 nm
Light Source:Miniature xenon flashlamp with interference filter
Material:Housing: Stainless steel 316Ti (1.4571)
Measurement Method:UV fluorescence method for polycyclic aromatic hydrocarbons (PAH)
Pressure Limit:Max. 30 bar or 435 psia (measurement probe)
Protection Class:IP 68
Range:High Measuring Range 0.1 to 15 ppm (mg/L) (oil) or 0 - 500 ppb (µg/L) (PAH) based on calibration standards
Reproducibility:2.5 % of measured value at constant temperature
Resolution:0.1 ppb (µg/L) (PAH) in the lowest measuring range
Response Time T90:10 s
Sample Temperature:1 °C to 40 °C (33.8 °F to 104 °F)
Storage Conditions:-40 °C to 60 °C
Weight:Stainless Steel: 6.17 lbs. (2.8 kg)