Using advanced ultraviolet (UV) absorption technology, Hach NITRATAX sc UV Nitrate Sensors offer unprecedented simplicity, accuracy, and economy in nitrate analysis. By continuously measuring the UV light absorbed by nitrates, NITRATAX sc sensors eliminate the need for reagents, sample conditioning, and frequent calibrations.
NITRATAX sc sensors rely on the principle that molecular bonds absorb ultraviolet (UV) light. In this case, nitrate (NO3) and nitrite (NO2) absorb UV light. As the concentration of nitrate or nitrite increases, UV absorption also increases. A built-in photometer measures the primary beam, while a second beam of UV light provides a reference standard and corrects for interference caused by turbidity and organic matter. Results are reported on the controller unit's alphanumeric display in mg/L of nitrate (NO3) or nitratenitrogen (NO3-N).
NITRATAX plus sc Process sensor for continuous measurement in drinking water, wastewater, or activated sludge. Turbidity compensation using reference measurement. NITRATAX eco sc low cost sensor for measurement, especially for sewage treatment plants with intermittent aeration technology. Turbidity compensation using reference measurement. NITRATAX clear sc Process sensor for continuous measurement in clean water such as drinking water or wastewater effluent.
With the sensor submerged in the sample stream, the detector windows are automatically cleaned by a built-in wiper that eliminate surface films or particles that can diminish accuracy.
An optional Bypass Panel or flow-through sample cell is available for clean water applications when direct immersion inOptional Bypass Panel a sample stream is impractical. The Bypass Panel uses the same probes as the immersion sensors for greater flexibility.
Nitrate probe, controller, flow through cell, standard solution, user manual
Accuracy:± 5 % of reading or ±0.5 mg/L, whichever is greater
Ambient Temperature:2 to 40 °C (36 to 100° F)
Cable Length:10 m (32.8 ft.) fixed cable at sensor
Dimensions (D x L):2.95 in x 12.72 in (75 mm x 323 mm)
Dimensions (H x W x D):8.3 in x 19.7 in x 6.3 in (210 mm x 500 mm x 160 mm)
Length:12.7 in ( 323 mm )
Maintenance Requirements:typical 1 hour/month
Measuring Interval:5 to 30 min (fixed values selectable)
Measuring Principle:Reagent-free UV absorption with patented 2-beam technique
Operating Pressure:0.5 bar (7.2 psi), maximum
Operating Temperature Range:2 to 40 °C
Path Length:5 mm
Pressure Limit:≤ 0.5 bar
Range:1.0 to 20.0 mg/L NO2+3–N
Resolution:0.1 mg/L NO2+3–N
Response Time:(T100) 5 min
Sample Flow Rate:0.5 - 10 L/h sample, connection: Tube ID 4 mm/AD 6 mm, Temp: 2-40°C,
Service Interval:6 months or as experience dictates
Verification:Using standard solutions
Weight:7.94 lbs. (3.6 kg) without sensor