介電常數(shù)及介質(zhì)損耗測(cè)試儀
編號(hào):GD201310250002
規(guī)格:GCSTD-A
高頻測(cè)量:10KHZ--50MHZ
介電常數(shù)測(cè)試儀及介質(zhì)損耗測(cè)試儀
本產(chǎn)品是依據(jù)版本國(guó)家標(biāo)準(zhǔn):GBT 1409-2006測(cè)量電氣絕緣材料在工頻、音頻、高頻(包括米波波長(zhǎng)在內(nèi))下電容率和介質(zhì)損耗因數(shù)的推薦方法
一、介電常數(shù)測(cè)試儀及介質(zhì)損耗測(cè)試儀概述
介質(zhì)損耗和介電常數(shù)是各種電瓷、裝置瓷、電容器等陶瓷,還有復(fù)合材料等的一項(xiàng)重要的物理性質(zhì),通過(guò)測(cè)定介質(zhì)損耗角正切tanδ及介電常數(shù)(ε),可進(jìn)一步了解影響介質(zhì)損耗和介電常數(shù)的各種因素,為提高材料的性能提供依據(jù);儀器的基本原理是采用高頻諧振法,并提供了,通用、多用途、多量程的阻抗測(cè)試。它以單片計(jì)算機(jī)作為儀器的控制,測(cè)量核心采用了頻率數(shù)字鎖定,標(biāo)準(zhǔn)頻率測(cè)試點(diǎn)自動(dòng)設(shè)定,諧振點(diǎn)自動(dòng)搜索,Q值量程自動(dòng)轉(zhuǎn)換,數(shù)值顯示等新技術(shù),改進(jìn)了調(diào)諧回路,使得調(diào)諧測(cè)試回路的殘余電感減至zui低,并保留了原Q表中自動(dòng)穩(wěn)幅等技術(shù),使得新儀器在使用時(shí)更為方便,測(cè)量值更為精確。儀器能在較高的測(cè)試頻率條件下,測(cè)量高頻電感或諧振回路的Q值,電感器的電感量和分布電容量,電容器的電容量和損耗角正切值,電工材料的高頻介質(zhì)損耗,高頻回路有效并聯(lián)及串聯(lián)電阻,傳輸線的特性阻抗等。
該儀器用于科研機(jī)關(guān)、學(xué)校、工廠等單位對(duì)無(wú)機(jī)非金屬新材料性能的應(yīng)用研究。
二、介電常數(shù)測(cè)試儀及介質(zhì)損耗測(cè)試儀的技術(shù)指標(biāo)
1.Q值測(cè)量
a.Q值測(cè)量范圍:2~1023。
b.Q值量程分檔:30、100、300、1000、自動(dòng)換檔或手動(dòng)換檔。
c.標(biāo)稱誤差
項(xiàng) 目 GCSTD-A
頻率范圍 20kHz~50MHz;
固有誤差 ≤5%±滿度值的2%;
工作誤差 ≤7%±滿度值的2%;
頻率范圍 10MHz~60MHz;
固有誤差 ≤6%±滿度值的2%;
工作誤差 ≤8%±滿度值的2%。
2.電感測(cè)量范圍:14.5nH~8.14H
3.電容測(cè)量:1~ 460
項(xiàng) 目 GCSTD-A
直接測(cè)量范圍 1~460pF
主電容調(diào)節(jié)范圍
準(zhǔn)確度 30~500pF
150pF以下±1.5pF;
150pF以上±1%
注:大于直接測(cè)量范圍的電容測(cè)量見(jiàn)使用規(guī)則
4.信號(hào)源頻率覆蓋范圍
項(xiàng) 目 GCSTD-A
頻率范圍 10kHz~50MHz
頻率分段
(虛擬) 10~99.9999kHz
100~999.999kHz
1~9.99999MHz
10~60MHz
頻率指示誤差 3×10-5±1個(gè)字
5.Q合格指示預(yù)置功能
預(yù)置范圍:5~1000。
6.Q表正常工作條件
a. 環(huán)境溫度:0℃~+40℃;
b.相對(duì)濕度:<80%;
c.電源:220V±22V,50Hz±2.5Hz。
7.其他
a.消耗功率:約25W;
b.凈重:約7kg;
c. 外型尺寸:(l×b×h)mm:380×132×280。
8.產(chǎn)品配置:
a.測(cè)試主機(jī)一臺(tái);
b.電感9只;
c.夾具一套
9、樣品尺寸:直徑為Φ50mm/Φ38mm 厚度為0.5-15MM
滿足客戶需要,現(xiàn)整理了英文資料如下:
GCSTD-A
The dielectric constant and dielectric loss tester tester
Meet the standards: GBT 1409-2006 measurement of electrical insulating materials at power, audio, high-frequency (including wavelength, VHF)recommended methods of permittivity and dielectric loss factor
One, overview
The dielectric loss and dielectric constant is all kinds of porcelain,porcelain, ceramic capacitors and other devices, and composite materialsis an important physical properties, by measuring the dielectric loss tangentof Tan δ and Permittivity (epsilon), can further understand the factors influencing the dielectric loss and dielectric constant, and provide the basis for improve the properties of materials; the basic principle of instrument is using high frequency resonance method, and provides a general, impedance test, multi use, multi range. It uses single-chip computer as core control measuring instrument, using the frequency locking, frequency standard test point automatic setting, resonant point automatic search, Q value range automatic conversion, numerical display and other new technology, improved the tuning circuit, the tuning circuit for testing the residual inductance is reduced to the minimum, and retains the original Q table in the automatic amplitude stabilizer technology, makes the new instrument in the use of more convenient, more accurate measurement. Instrument in high frequency test conditions, measurement of high frequency inductor and resonant loop of the Q value, inductance and distributed capacitance of inductors, capacitance and loss tangent of dielectric loss, high frequency electrical materials, high frequency circuit in parallel and series resistance, the characteristic impedance of the transmission line
===================================
Research and application of scientific research institutions, schools,factories and other units of the new material properties of inorganic nonmetal for the instrument.
Technical indicators two, instrument
The measurement of 1.Q value
The a.Q value range: 2 ~ 1023.
The b.Q value range shift: 30, 100, 300, 1000, automatic or manual shift.
C. error
Item GCSTD-A
The frequency range of 20kHz to 50MHz;
The inherent error ≤ 5% ± full scale value 2%;
Work with error less than 7% ± full scale value 2%;
The frequency range of 10MHz to 60MHz;
The inherent error ≤ 6% ± full scale value 2%;
Work with error less than 8% ± full scale value 2%.
The 2 inductance measurement range: 14.5nH~8.14H
The 3 capacitance measurement: 1~ 460
Item GCSTD-A
Direct measurement of the range of 1 ~ 460pF
The main capacitor adjustment range
The accuracy of 30 ~ 500pF
The following 150pF + 1.5pF;
150pF or + 1%
Note: greater than direct measurement range of capacitance measurementsee using rules
Frequency of the 4 signal coverage
Item GCSTD-A
The frequency range of 10kHz to 50MHz
The frequency segment
(virtual) 10 ~ 99.9999kHz
100 - 999.999kHz
1 - 9.99999MHz
10 - 60MHz
Frequency indication error of 3 × 10-5 ± 1 words
5.Q conformity indicating preset function
The preset range: 5 ~ 1000.
Table 6.Q normal working conditions
A. environment temperature: 0 ℃ ~ +40 ℃;
Relative humidity: <80% b.;
C. power supply: 220V ± 22V, 50Hz ± 2.5Hz.
7 other
A. power consumption: about 25W;
B. net weight: about 7kg;
C. size: (L * b * h) mm:380 x 132 x 280.
8 the product configuration:
A. a test mainframe;
B. inductor 9;
A C. clamp
9, the sample size: diameter is phi 50mm/ Phi 38mm thickness of 0.5-15MM